7 Replacing components
Service Manual Sinius / Sinius CS / Sinius TS
7.3 Important information when replacing boards
båÖäáëÜ
7.3.3
Replacing the HSA board in the patient chair
The reference of the travel track is stored in the HSA board.
✔
You have replaced the HSA board.
1. Perform a reference travel. For information on how to do this, see
Referencing the drives [ → 35].
2. Check the date and time in the Setup dialog box.
7.3.4
NSK board in the patient chair
The reference of the headrest is stored in the NSK board.
✔
You have replaced the NSK board.
➢ Perform a reference travel. For information on how to do this, see
Referencing the drives [ → 35].
More information can be found under: Removing the cover of the
motorized headrest on the patient chair [ → 106]-> Cable routing when
fitting cover parts.