4 Technical data
Service Manual Sinius / Sinius CS / Sinius TS
4.1 Operating, transport and storage conditions
båÖäáëÜ
4.1
Operating, transport and storage conditions
Operating mode:
Continuous operation with intermittent
loading corresponding to the dental
mode of working.
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.
Transport and storage
conditions:
Temperature: -40 °C – +70 °C
(-40 °F – 158 °F)
Relative humidity: 10% - 95%
Air pressure: 500 hPa – 1060 hPa
Operating conditions:
Ambient temperature: 10 °C – 40 °C (50
°F – 104 °F)
Relative humidity: 30% - 85%
no condensation
Air pressure: 700 hPa – 1060 hPa
Installation location:
≤ 3000 m above sea level
This treatment center is not suitable for
operation in areas subject to explosion
hazards.
Pollution degree:
2 acc. to IEC 60664-1
-40
+70
1060
500
95
10
+10
+40
1060
700
85
30
Local pressure (min./
max.):For details see page
Requirements of the supply
media
Air: 5.5 / 7.5 bar (79.8 / 108.8 psi)
Water: 2.5 / 6 bar (36.3 / 87 psi)
Suction air: p
u
min. 0.12 bar (1.7 psi),
max. 0.18 bar (2.6 psi)
≥ 500 Nl/min
Water flow:
For details see page
Requirements of the supply
media
Water absorption max. 3 l/min
Residual water max. 3 l/min