10 Biofilm removal
Service Manual Sinius / Sinius CS / Sinius TS
10.3 Execution
båÖäáëÜ
Rinsing / cleaning solution 2
1. Rinse all instrument hoses carrying water for approx. 10 seconds by
activating the relevant instrument multiple times. Switching the water
flow on and off supports the process for removing deposits.
2. Activate the tumbler filler 1x.
3. Activate the flushing 1x.
4. Activate the suction hose cleaning 1x (if present).
5. Rinse the external device connection (if present) by opening the valve
until water flows out of the hose. Close the valve again.
6. Let cleaning solution 2 take effect in accordance with the instructions
from the manufacturer of the cleaning agent.
7. Complete steps 1-6 until the tanks are empty of all solution (including
the mixing and rinsing tank).
8. Switch off the pump on the biofilm removing system.
IMPORTANT
Collect the colored cleaning liquid as applicable in a suitable container
for subsequent inspection.
IMPORTANT
The solution in the collection container must be clear following the
rinsing process. If the solution is still cloudy, fill the tanks with a new
cleaning solution 2 again and complete cleaning steps once again. The
cleaning liquid can then be removed via the cuspidor.