10 Biofilm removal
Service Manual Sinius / Sinius CS / Sinius TS
10.4 Re-processing
båÖäáëÜ
5. Insert the new filter at the removed points.
6. Install the corner valve with the gaskets onto the sanitary hose and
open it.
7. Rinse all instrument hoses carrying water for approx. 30 seconds by
activating multiple times.
8. Activate the tumbler filler 1x.
9. Activate the flushing 1x.
10. Activate the suction hose cleaning 1x (if present).
11. Rinse the external device connection (if present) by opening the valve
until water flows out of the hose. Close the valve again.
12. Repeat steps 6-10 until the only water coming out of all channels
carrying water is clear.