63 69 149 D3561
D3561
.
076.01.07
.
02 02.2017
50
Dentsply Sirona
6 Overview of modules and boards
Service Manual Sinius / Sinius CS / Sinius TS
6.2 Patient chair
båÖäáëÜ
V306
BOOT
(yellow)
Lights up after switching on and remains lit until completion
of the boot process (approximately 60 seconds)
● The application cannot be launched
– Reset the board
– The boot LED goes out briefly and then lights up
again
– All network LEDs light up
● The boot LED goes out briefly and then lights up again
– The run LED starts flashing
– The application has been launched
– The download manager is running as long as the
boot LED is lit
Mode
(yellow)
Off: Normal
On: Installation mode without dentist element
Flashing: Active production test mode
V309
T26
(yellow)
Reserved for internal debugging options
V310
T31
(yellow)
Reserved for internal debugging options
V405
HR_stop
(yellow)
ON: Movement lock due to active instrument
V410
Chair stop
(yellow)
ON: Movement lock due to active instrument
V604
LED suction
(yellow)
ON: Relay switched
V620
LED bell
(yellow)
ON: Relay switched
V621
LED AUX
(yellow)
ON: Relay switched
V801
FS active
(yellow)
ON: Foot control active
Collision
(yellow)
Heavy data traffic
V904
Full duplex
(yellow)
ON: For full duplex
OFF: For half duplex
V903
10MB
(yellow)
ON: 10Mbit
V902
100 MB
(yellow)
ON: 100 Mbit
V905
Link
(yellow)
ON: Connected
Flashing: LAN data transmission
SCOM
(green)
Speed-Com status
ON: HAP/HSA data connection OK
Flashing: Connection being established or error (cable
connection present)
OFF: No data connection present (disruption in the line)
V1008
Move lift
(yellow)
Movement release for raising chair
V1003
Move backrest
(yellow)
Movement release for backrest
V1009
Move chair
(yellow)
Movement release for chair
HUB_OC
(yellow)
Components on the board