Dentsply Sirona
6 Overview of modules and boards
Service Manual Sinius / Sinius CS / Sinius TS
6.4 Water unit
båÖäáëÜ
CUPFILL
(yellow)
ON: Morita > switch on MV34 (tumbler filling)
CHAIR_STOP
(yellow)
ON: Movement lock due to active instrument
MV3
(yellow)
ON: Metering valve for water sterilizing agent switched on
SIR_ON
(yellow)
ON: Examination light switched on
ROE_ON
(yellow)
ON: X-ray viewer activation pulse
ROE_OFF
(yellow)
ON: X-ray viewer deactivation pulse
MS_ON
(yellow)
ON: Sepa motor switched on
MA_ON
(yellow)
ON: Amalgam separator switched on
V1_LED
(yellow)
ON: Light barrier signal while amalgam separator is running
MV40.1
(yellow)
ON: Right-hand chamber of the water pump active
MV40.2
(yellow)
ON: Left-hand chamber of the water pump active.
Alternating flashing of MV40.1 and MV40.2 LED when water
is being removed
MV41
(yellow)
ON: Fill valve for rinsing tank switched on
MV44
(yellow)
ON: Hydrocolloid valve switched on
MV1
(yellow)
ON: Fill valve for mixing tank switched on
MV34
(yellow)
ON: Fill valve for tumbler switched on
MV33
(yellow)
ON: Place selection valve for suction system switched on
AA95
(yellow)
ON: Amalgam rotor fill level >95%
AA100
(yellow)
ON: Amalgam rotor fill level 100%
MV37
(yellow)
ON: Waste drain flap to the amalgam separator open
MV35
(yellow)
ON: Flushing switched on
Connector
X1
> L207 > X1 power supply
X2
> Amalgam motor (amalgam separator)
X3
> L206 > X3 power supply
X4
> Speed sensor (amalgam separator)
X5
> L400A > X2 HSA
X7
> AW safety switch (amalgam separator)
X8
> L210 > Water detector (M37 and DN30)
X9
> L219 > M35 flushing
X10
> DNA float switch
X11
> L221 MV34 cup fill
X12
> MV33 automatic separator suction valve
X13
> DS30 automatic separator level switch
X14
> L245 > Heater
X15
> X-ray viewer
X18
> DNA cuspidor bowl valve
X20
> L418 > S13 cuspidor bowl
Components on the board