63 69 149 D3561
D3561
.
076.01.07
.
02 02.2017
82
Dentsply Sirona
6 Overview of modules and boards
Service Manual Sinius / Sinius CS / Sinius TS
6.4 Water unit
båÖäáëÜ
Function of the board
The board controls all functions listed below:
● Suction hose purging
● Chemical disinfection system
● Hydropneumatic pump with flowrate measurement using Hall
sensors
● Standard tumbler filling with heater
● Automatic tumbler filling
● Automatic separation
● Amalgam separator with water alarm
● Wet suction with cleaning button
● Flushing
● X-ray image viewer on the tray
● Hydrocolloid solenoid valve
● LEDview / LEDview Plus examination lights
● Switching off the central power supply
● Acoustic signal sensor for amalgam alarm, programming
confirmation, etc.
● Fill level detection using magnetically actuated reed contacts:
– Mixing tank with upper (DS31.1) and lower (DS31.2)
– fill level sensor
● Disinfection tank with lower fill level sensor (DS41)