10 Biofilm removal
Dentsply Sirona
10.4 Re-processing
Service Manual Sinius / Sinius CS / Sinius TS
10.4
Re-processing
1. Reduce the water pressure in the treatment center, by removing the
Sprayvit and operating the water button until no more water emerges.
2. Switch the treatment center off.
3. Reduce the air pressure in the treatment center by operating the air
button.
4. Remove the biofilm removal device.
WARNING
Drops of the cleaning solution contain acids / alkalines and may affect
surrounding parts.
Carefully loosen the adapter from the sanitary hose.