BENXSPSDFUWEB / Rev.09 / 10-2018
39/60
Step-by-Step Instructions
6.11.2 Lubricating the Contra-angle
• Lubricate the contra-angle only with a dedicated spray.
• Lubricate after each use and before sterilization.
1.
Screw the spray nozzle onto the spray with approx. 10 turns.
Contra-angle
F type Spray Nozzle
2.
Insert the spray nozzle into the rear part of the contra-angle and lubricate
for 2-3 seconds until oil comes out from the contra-angle head.
3.
Before attaching the lubricated contra-angle to the motor handpiece, wipe
off excess oil. Place it on its end or lean it at an angle for gravity draining.
Mount it after excess oil has been drained.
WARNING
Do not lubricate the motor handpiece.
CAUTION
• Hold the contra-angle securely to prevent it from detaching with
the pressure of the spray.
• Never use a spray can upside down. Only spray gas, and no oil,
would be sprayed.