BENXSPSDFUWEB / Rev.09 / 10-2018
47/60
Error Code
Error
code
Error
Cause
Check/Remedy
During
rotation of
the motor
handpiece
E-00
Self-Check
Malfunction of circuit.
Contact your
distributor.
E-01
Over current
The motor handpiece
is locked (at the time
of the auto reverse
mode).
Remove load.
The motor handpiece
cord has shorted.
Contact your
distributor.
E-02
Over voltage
Malfunction of circuit.
Contact your
distributor.
E-03
Motor
handpiece
sensor
The motor handpiece
cord is disconnected.
Connect the motor
handpiece cord
correctly.
Faulty sensor
(Hall IC). Severed
cord (signal line).
Contact your
distributor.