Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should ensure that it is used in such an
environment.
Immunity test
IEC 60601-1-2 test /
compliance level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 Vrms from 150 kHz to
80 MHz
6 Vrms in the ISM
bands
Portable and mobile RF
communications equipment should be
used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from that equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2
√
P
d = 1,2
√
P 80 MHz to 800 MHz
d = 2,3
√
P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should
be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC 61000-4-3
3 V/m from 80 MHz to
2.7 GHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
9 V/m in the bands
710 MHz, 745 MHz,
780 MHz, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m in the band
385 MHz
28 V/m in the bands
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz