B EN XS00 DFU WEB / Rev.08 / 01-2018 (Old ZF 190270.EN)
33/48
Changing batteries
The control unit, the micromotor (handpiece), including all wires may be
wiped with a cloth slightly dampened with water or non-alcoholic
disinfectant.
10
CHANGING BATTERIES
X-Smart
®
uses rechargeable batteries. They can be recharged 300 -
500 times, depending on the conditions of the motor handpiece. If the
operating time or recharging time becomes shorter or the rotation power
becomes weaker, although the “memory effect” described in
“
NOTICE” is not applicable, the life of the batteries is suspect. In
such a case, replace batteries (refer to the “Changing Batteries”). When
replacing, be sure to observe the following “
CAUTIONS ON
CHANGING BATTERIES”. Note that Dentsply Sirona shall not be held
liable for any malfunction or failure resulting from your not following the
“
CAUTIONS ON CHANGING BATTERIES”
.
CAUTION
• Do not apply high-temperature sterilization methods such as
autoclave sterilization to any part other than the contra angle
head.
• Skip dry cycle if the temperature could exceed 135°C (275°F).
• Since the bottom tray inside the chamber of the autoclave is close
to a heater and the temperature in that place may locally exceed
135°C (275°F), place the head on the central or upper tray.
CAUTION ON CHANGING BATTERIES
• Do not open any part other than the battery cover.
• Be sure to purchase and use only the recommended batteries
(part reference A1007 000 00 100). Otherwise, batteries may
cause damage, fluid leakage or explode.
• Do not work with wet hands. This may cause trouble due to short-
circuiting of batteries and moisture infiltrating this product.