Troubleshooting
38/48
B EN XS00 DFU WEB / Rev.08 / 01-2018 (Old ZF 190270.EN)
AC IN lamp does not
light.
The AC adapter is not
connected.
Check the
connection.
The plug of the AC
adapter is not inserted
into the outlet, or there is
no electricity in the outlet.
Check the
connection.
The internal fuse has
burnt.
Contact your dealer.
The AC adapter fuse has
burnt.
Contact your dealer.
The battery charger
does not work (the
CHARGE lamp does
not light).
No batteries are inserted.
Insert batteries.
The batteries are fully
charged or in a state near
full charge.
No problem.
The temperature of
batteries is low.
If the temperature of
batteries is less than
0°C (32°F), the
batteries are not
rechargeable. Charge
the batteries in a
warm room (be
careful about
moisture
condensation).
The temperature of
batteries is high.
It is normal that the
batteries become a
little bit warm right
after charging. If the
batteries are hot
under normal
operating conditions,
not right after
charging, there may
be an abnormality.
Contact your dealer.
An error code is
displayed.
See
11 Error Code
.
Problem
Cause
Solution