OPERATING INSTRUCTIONS
•
Turn the furnace on
•
Insert muffle door
•
Turn temperature control dial to 650°C (1200°F)
•
Turn the rate control knob to the midpoint between "S" and "F"
•
After 15 minutes, adjust the temperature control knob to 1000°C (1832°F) and the rate to maximum ("F").
•
Allow furnace to maintain temperature for 10 minutes and readjust the temperature to the desired temperature.
After the Miniglaze has been calibrated and the initial heating procedure has been completed it is ready to process
dental porcelain.
Staining and Glazing
Before each operation, the furnace should be brought up to 540°C (1000°F). Set the temperature dial at the desired
temperature and adjust the temperature rise to the desired level. Using a pair of furnace tongs place the work in the
muffle and replace the door. The tone alarm or light will signal that the set temperature has been reached.