Operation Guidelines - Quick Start
1
Unit Set-Up
a. Connect the power cord to the back
of the units and to an outlet.
b. Turn the power switch found on
the back of the unit to the on position.
c.
LED
displays the last timer setting. If the timer
needs to be adjusted, touch
Time
button.
After first use, timer will remain set to same
time unless changed.
d. Select desired time setting by touching
Time button.
Please consult your instrument cleaner instructions
as well as your instrument directions for use for
proper automated cleaning.
Prolonged cleaning time may result in:
a. Loosening of the screws if used.
b. If items are plated and have existing
cracks, the cracks may become more
pronounced.
c. Coating with preexisting damages
may experience more peeling.
Degas Procedure
a. Add warm water to below the max
fill line.
b. Add Resurge Instrument Cleaning
Solution per the instructions for use.
c. Touch the degas key and allow it to
operate for 3 minutes.
It is best to remove the gases from the
solution to maximize the cavitation
energy. If you do not degas, the
cavitation energy can be blocked by the
gases making the cavitation energy less.