Operation

Set the fan speed switch on the rear of the unit to the LOW, 
MED, or HIGH position.  Higher airflow will result in faster 
discharge times.  Position the unit so that the maximum 
airflow is directed at the items or area to be neutralized.  
Turn the unit ON.

When the unit is first turned on, it will conduct a self-test.  
The audible alarm will sound and then the LED will cycle 
through RED, YELLOW, and GREEN.  The LED will remain 
GREEN during normal operation.

BALANCE (OFFSET VOLTAGE) ADJUSTMENT

The Chargebuster Overhead Ionizer is an auto-balancing 
unit.  However, tuning or manual adjustment can be 
accomplished by inserting a small screwdriver or trimmer 
adjustment tool into the balance (offset voltage) adjustment 
hole located on the front of the unit.  To increase the output 
in a positive direction, turn the potentiometer in a clockwise 
direction.  Conversely, to increase the output in a negative 
direction, turn the potentiometer in a counter-clockwise 
direction.  Once the desired balance point has been 
achieved, the alarm reset button must be pressed to set the 
balance point.  

MAINTENANCE / ALARMS

If the supply voltage drops from 120 Volts to below 85 Volts, 
the unit will shut down, the audible alarm will beep, and the 
LED will blink RED.  The unit will automatically reset when 
the minimum voltage is restored.

Under normal conditions the ionizer will attract dirt and dust 
(especially on the emitter electrodes).  To maintain optimum 
neutralization efficiency and operation, cleaning should be 
performed on a regular basis. 

When the unit enters shutdown mode, ionization will be 
stopped, the LED on the front of the unit will change to a 
steady RED, and the audible alarm will sound continuously.  
If the ionizer enters shutdown mode, it must be turned OFF 
and then back ON to reset the unit.

NOTE:  The AC power cord MUST be disconnected 
before the unit is disassembled for maintenance.

  The 

emitter electrodes should be cleaned using the included 
cleaners or a swab damp with isopropyl alcohol.  First, 
turn the unit OFF and unplug the power cord.  Then 
remove each fan grill located on the top-side of the ionizer 
by disengaging all 4 screws per grill.  Clean all emitter 
electrodes.  After cleaning the emitter electrodes, re-attach 
all the removed grills.  Plug in the power cord and turn the 
unit back ON.

The emitter electrodes should not require replacement 
during the life of the unit with normal handling.  
Replacement emitter electrodes can be ordered if 
necessary.

The best practice would be to verify the balance (offset 
voltage) of the unit with a charged plate monitor after 
cleaning.

NEUTRALIZATION (DISCHARGE) TIMES

The comparative efficiency of overhead ionizers is 
determined by a standard test published by ANSI/ESD 
S3.1.  Typical positive and negative discharge times (in 
seconds from 1000 volts to 100 volts and from -1000 volts 
to -100 volts) measured using this standard are shown 
below.  The performance of the ionizer was measured with 
the unit positioned as shown, with the fan speed on high, 
and without a filter.

The typical discharge times measured using this standard 
for the ionizer are shown in the following diagram. They are 
not guaranteed maximum discharge times.

TB-3031 Page 2 of 4

© 2010 DESCO INDUSTRIES INC.

Employee Owned

DESCO WEST

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449 

DESCO EAST

 - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: 

Desco.com

Figure 5.  Neutralization (discharge) times - 2 fan ionizer

Figure 3. 110 volt jumper setting

Figure 4. 220 volt jumper setting

Summary of Contents for Chargebuster

Page 1: ...nce Install the unit in the desired location making sure that the airflow will not be restricted Be sure the OFF ON switch located on the rear of the unit is in the OFF position Plug the power cord into the unit and then into the appropriate AC power source TB 3031 May 2010 Page 1 of 4 Figure 2 Desco 60640 Chargebuster Overhead Ionizer Description The Chargebuster overhead ionizers are steady stat...

Page 2: ... alarm will sound continuously If the ionizer enters shutdown mode it must be turned OFF and then back ON to reset the unit NOTE The AC power cord MUST be disconnected before the unit is disassembled for maintenance The emitter electrodes should be cleaned using the included cleaners or a swab damp with isopropyl alcohol First turn the unit OFF and unplug the power cord Then remove each fan grill ...

Page 3: ...cation of ESD Protective Equipment and Materials The most accurate tool to use for this is a Charged Plate Monitor Alternatively a portable battery operated Ionization Test Kit can be used Per ESD TR 53 ANNEX A Test Frequency The objective of the periodic test procedures listed in this document is to identify if significant changes in ESD equipment and materials performance have occurred over time...

Page 4: ...dentify each ionizer and setup a compliance Verification Maintenance Calibration schedule If the ionizers all test good the data can justify lengthening the calibration period If ionizers require adjustment the calibration period should be shortened Although ESD TR53 does not advise a test frequency JESDD625 A Revision of EIA 625 recommends ionizers be tested semiannually noting to use S3 1 except...

Reviews: