14-26 SDI-12 Operations
XL™ Series
Example of a "Write User Units Offset" command:
Command
Response
Time
Values
Description
---------------
-----------------
-------
--------
-----------------
"aXWO12.34!"
"a0021<cr><lf>"
2 sec
1
Write user units offset
coefficient
The Write User Units Offset command loads the "User Units Offset" coefficient term. The new
value is stored in non-volatile EEPROM within the sensor. A copy of this value is placed in the
sensor data buffer for verification. It can be viewed by using the "D" command. To verify the
"User Units Offset" at any other time, use the "Read User Units Offset" command.
Command
Response
Description
-------------
------------------------- ---------------------------
"aXWOddd!"
"atttn<cr><lf>"
Write User Units Offset coefficient
Where:
a
Is the sensor address ("0" to "9", "A" to "Z", or "*").
XWO Are upper-case ASCII characters.
ddd
Is the new User Units Offset value. The input format is very flexible. Some
examples are shown below.
20.095
0.195
7.984E+10
167.824E5
005.9357E-7
500
ttt
Is a three digit integer (000-999) specifying the maximum time, in seconds, the
sensor will take to complete the command and have data available in its buffer.
n
Is a single digit integer (0-9) specifying the number of values that will be placed in
the data buffer.
Summary of Contents for H-350XL
Page 1: ...XL Series Models H 350XL H 500XL H 510XL H 522 H 522 Plus Owner s Manual ...
Page 30: ...2 16 Hardware Options and Installation XL Series ...
Page 42: ...3 12 Using The Built In Keypad Display XL Series ...
Page 46: ...4 4 PC Menu Interface Operation XL Series ...
Page 58: ...5 12 Miscellaneous System Setup XL Series ...
Page 66: ...6 8 Serial Port Options XL Series ...
Page 74: ...7 8 System Configuration Files XL Series ...
Page 84: ...8 10 System Status Menus Options XL Series ...
Page 88: ...9 4 Scanning Options XL Series ...
Page 156: ...15 12 Logging Options XL Series ...
Page 198: ...16 42 Operation with the GOES Radio XL Series ...
Page 216: ...19 6 Alarm Call Out XL Series ...
Page 234: ...21 8 Operation with the H 355 Smart Gas System XL Series ...
Page 246: ...22 12 Functions XL Series ...