XL™ Series
Scanning Options 9-3
PC Interface View Built In Display Screen
R - Scan Rate: HH:MM:SS
Scan Rate [HH:MM:SS]
This screen displays the current scan rate. The scan rate can be set in the range of 1 second to 24
hours. If the scan rate is set faster than the time the XL™ takes to make one scan then the actual
scan rate will be a multiple of the entered value. For example a 30 second scan rate will
automatically turn into a 1 minute scan rate if the measurement process takes 35 seconds to
complete.
PC Interface View Built In Display Screen
F - Start a New File:
Not Available In Built In Menu
This option causes the unit to start a new data file on the next scan without having to turn
scanning off. Using this option at the end of a site visit will cause all the data from that point on
to be in a new file so that on the next visit all the data in the file will be that collected between
the visits.
Summary of Contents for H-350XL
Page 1: ...XL Series Models H 350XL H 500XL H 510XL H 522 H 522 Plus Owner s Manual ...
Page 30: ...2 16 Hardware Options and Installation XL Series ...
Page 42: ...3 12 Using The Built In Keypad Display XL Series ...
Page 46: ...4 4 PC Menu Interface Operation XL Series ...
Page 58: ...5 12 Miscellaneous System Setup XL Series ...
Page 66: ...6 8 Serial Port Options XL Series ...
Page 74: ...7 8 System Configuration Files XL Series ...
Page 84: ...8 10 System Status Menus Options XL Series ...
Page 88: ...9 4 Scanning Options XL Series ...
Page 156: ...15 12 Logging Options XL Series ...
Page 198: ...16 42 Operation with the GOES Radio XL Series ...
Page 216: ...19 6 Alarm Call Out XL Series ...
Page 234: ...21 8 Operation with the H 355 Smart Gas System XL Series ...
Page 246: ...22 12 Functions XL Series ...