Pasteurization Procedure

High Level Disinfection

1. Create  a circulating water bath filled with sterile purified water (PURW) to 70°C (158°F).
2. Immerse the mask in the water bath and expose for 30 minutes.
3. Remove the mask from the water bath and air dry

.

1. Prepare a disinfectant compatible with Polycarbonate and Silicone according to

the manufacturer’s recommendations.

2. Immerse the mask completely, filling all areas and eliminating air pockets.
3. Following the disinfection exposure, thoroughly rinse the mask by immersing it 

completely in a large volume of purified water. Allow the mask to remain in contact 
with the water for a minimum of one minute.

4. Remove the mask and repeat the rinse procedure a minimum of two additional times

using fresh, purified water for each rinse.

5. After rinsing, allow the mask to completely air dry before use.

NOTE:

Follow all manufacturer’s instructions and guidelines for product use.

Note: Velcro

®

patches may release from the FlexSet forehead pad with repeated disinfection.

(Replacement Forehead Pad: p/n 9353D-604)

DeVilbiss Mask Cycle Tolerances

DeVilbiss Mask

Cleaning

Pasteurization

Serenity Silicone (Medium 9352D and Small 9352S)

26

10

Serenity Gel (Medium 9352G and Small 9352GS)

26

10

FlexSet Silicone (Medium 9354D and Small 9354S)

26

10

FlexSet Gel (Medium 9354G and Small 9354GS)

26

10