11
A-3655X
EN
SPECIFICATIONS
SPECIFICATIONS
Size
7.5” W x 4.0” H x 6.8” D
(19.1 cm x 10.2 cm x 17.3 cm)
Weight
4.1 lbs. (1.86 kg)
Electrical Requirements
3655DX & 3655TX
2.5 amps, 115 VAC (±10%) 60 Hz
3655IX
1.4 amps, 230 VAC (±10%) 50 Hz
3655KX & 3655PX
1.4 amps, 230 VAC (±10%) 60 Hz
Sound Level
56 dBA
Power Consumption
3655DX & 3655TX
140 watts maximum
3655IX, 3655KX & 3655PX
160 watts maximum
Maximum Compressor Pressure
35 psig or greater (241 kPa)*
Compressor Free Air Flow
9 lpm or greater*
Operating Temperature Range
+40° to +104°F (+5 to +40°C)
Operating Humidity
up to 95% non-condensing
Storage/Transport Temperature Range
-40° to +158°F (-40° to +70°C)
Storage/Transport Humidity
up to 95% non-condensing
Supplied Nebulizer Capacity
6 ml (cc)
Supplied Nebulizer Nebulization Rate
0.15 ml/min or greater*
Supplied Nebulizer MMAD
5 microns or less**
*Conditions may vary based on altitude above sea level, barometric pressure, and
temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are
no known risks associated with disposal of the equipment.
This product (models 3655DX & 3655TX) complies with the following electromagnetic
compatibility standard: IEC 60601-1-2:2nd Edition
3655DX & 3655TX Equipment Classifications:
Protection against electric shock: ................................................................................. Class II
Degree of protection against electric shock: ................................................................Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or
with oxygen or nitrous oxide.
Mode of operation: Continuous operation
Summary of Contents for 3655DX