EN - 11
SE-DV6WM-1
EMC INFORMATION
(per IEC 60601-1-2:2007 Section 5.2.1.1)
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in this instruction guide.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT
(per IEC 60601-1-2:2007 Section 5.2.2.1 c)
Guidance and Manufacturer’s Declaration –
Electromagnetic Emissions
The DeVilbiss DV6WM Wireless Modem is intended for use in the
electromagnetic environment specified below. The customer or the
user of the DeVilbiss DV6WM Wireless Modem should assure that it
is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment –
Guidance
RF Emissions
CISPR 11
Group 2
This device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
This device is suitable for use in all
establishments including domestic,
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonics IEC
61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
(per IEC 60601-1-2:2007 Section 5.2.2.1 f)
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic Environment
– Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95 % dip in UT for 0.5
cycle
60 % dip in UT for 5
cycles
30 % dip in UT for 25
cycles
>95 % dip in UT for 5
seconds
>95 % dip in UT for
0.5 cycle
60 % dip in UT for 5
cycles
30 % dip in UT for 25
cycles
>95 % dip in UT for 5
seconds
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the DeVilbiss DV6WM
Wireless Modem requires
continued operation during power
mains interruptions, it is
recommended that DeVilbiss
DV6WM Wireless Modem be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.