Changing power supply
1
.
6
.
1
Removing power supply
• Remove cover (see point
1
.
2
)
• Loosen plug connection J
3
• Loosen countersunk screws on base of device.
• Remove cable ties from transformer.
• Lift power supply upwards and remove.
1
.
6
.
2
Fitting power supply
• Place power supply into device.
• Connect plug connection J
3
to the board.
• Attach cable loom to the transformer using cable ties.
• Fix power supply to base of device using countersunk screws.
• Connect push button box to the board.
• Fill chamber with distilled water up to the mark and fit cover.
• Fit nebuliser chamber.
• Switch on device.
• Adjust the device following the setting instructions.
• Remove push button box and housing cover.
• Fit housing cover (see point
1
.
3
)
• Check electrical safety according to VDE
0751
or IEC
601
-
1
• Trial run
1
.
7
Changing fan
1
.
7
.
1
Removing device fan
• Remove cover (see point
1
.
2
)
• Loosen screws on the device fan holder
• Unsolder cable from the board (J
13
)
• Change fan
1
.
7
.
2
Fitting device fan
• Fit fan and fan holder. Ensure flow direction is correct! Air is
drawn out of the power supply. Observe direction of the
arrow on the side of the fan.
• Shorten the cable and solder device fan into J
13
(see Fig.
2
).
Ensure polarity is correct!
• Connect push button box to the board.
• Fill chamber with distilled water up to the mark and fit cover.
• Fit nebuliser chamber.
• Switch on device.
• Adjust the device following the setting instructions.
• Remove push button box and housing cover.
• Fit housing cover (see point
1
.
3
)
• Check electrical safety according to VDE
0751
or IEC
601
-
1
• Trial run
1
.
7
.
3
Removing patient fan
• Remove cover (see point
1
.
2
)
• Loosen the device fan screws underneath the device.
• Unsolder cable from the board (J
5
)
• Change fan
1.7.4
Fitting patient fan
• Fit fan.
• Shorten the cable and solder device fan into J
5
(see Fig.
2
).
Ensure polarity is correct!
• Connect push button box to the board.
• Fill chamber with distilled water up to the mark and fit cover.
• Fit nebuliser chamber.
• Switch on device.
• Adjust the device following the setting instructions.
• Remove push button box and housing cover.
• Fit housing cover. (see point
1
.
3
)
• Check electrical safety according to VDE
0751
or IEC
601
-
1
• Trial run.