Cirona
TM
Deep Vein Thrombosis (DVT) Therapy Garment
1. Plug the pump into an electrical outlet but DO NOT SWITCH ON YET.
2. Remove the sleeve(s) from packaging.
3. The garments can be used on either foot depending on conditions.
4. Attach the tube set to the pump by connecting the male connectors into the pump connectors until a click
is heard. Pump connectors are located on the right side of the pump.
5. Attach the garments to the tube set using the snap lock connectors.
6. Foot Garment: Fully unfold and position the inflatable chamber against the base of the patient’s foot with
the tube set facing upward. Wrap the tab firmly over top of the foot. Pulling the remaining long tab around
the heel and firmly secure to the top of the foot.
(See reverse side for visual instruction.)
7. Switch on the device. Please refer to the device’s Operation Instructions for how to use the system.
01 Application Instructions
IFU37.0009 Rev C 20141111
•
Ensure that there are no blockages in the tubing and that connectors are firmly locked in place.
•
Regularly check the system while it is in operation and ensure the garment fits.
•
Garment should be removed if the patient experiences pain, numbness, or tingling.
•
Never put on or take off the garments when inflated as it may cause damage to the sleeves.
02 Cautions
The Cirona
TM
is a prescription-only device intended to be used preventatively to increase venous blood flow
in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during:
trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, urologic,
obstetric conditions and treatments.
The Cirona
TM
should NOT be used on patients with the following conditions:
1. Severe atherosclerosis or other ischemic vascular diseases
2. Suspected or known acute deep vein thrombosis
3. Severe congestive cardiac failure
4. Existing pulmonary edema
5. Extreme deformity of the limbs
6. Any local skin or tissue condition in which the garments would interfere:
•
Gangrene
•
Untreated or infected wounds
•
Recent skin graft
•
Dermatitis
7. Known presence of malignancy in the legs
8. Limb infections, including cellulitis, that have not received antibiotic coverage
9. Presence of lymphangiosarcoma
03 Intended Use
04 Contraindications
Non-Sterile* Garments. Intended for single patient use and cleaning is not recommended
Foot Garments
Model Name Size (Instep Circumference)
F-15 Small Foot Up to 13”
F-10
Standard Foot Up to 16”
© 2014 Devon Medical Products • 1100 First Avenue, Suite 202 • King of Prussia, PA 19406 • 866.446.0092 • info@devonmedicalproducts.com