19
20
Recommended separation distances between portable and mobile RF
communications equipment and the
model extriCARE™ 2400
.
The
model extriCARE™ 2400
is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of
the
model extriCARE™ 2400
can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
model extriCARE™ 2400
as
recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
Separation distance according to frequency of transmitter (m)
output power of
transmitter
(W)
0.01
0.1
1
10
100
150 kHz to 80 MHz
d = 1.2
√
P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2
√
P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3
√
P
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.