Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
mmol/L), and 96% (238 out of 248) of the YSI values were greater than
280 mg/dL (15.6 mmol/L).
Table 2. Number and Percentage of YSI Values When
CGM Readings are ‘Low’ or ‘High’
YSI mg/dL (mmol/L)
Total
CGM
Readings
CGM-YSI
pairs
< 55
(3.1)
< 60
(3.3)
< 70
(3.9)
< 80
(4.4)
> 80
(4.4)
‘LOW’
n
66
18
39
19
13
155
Cumulative
Percent
42%
54%
79%
92%
8%
YSI mg/dL (mmol/L)
Total
CGM
Readings
CGM-YSI
pairs
> 340
(18.9)
> 320
(17.8)
> 280
(15.6)
> 240
(13.3)
< 240
(13.3)
‘HIGH’
n
189
31
18
8
2
248
Cumulative
Percent
76%
89%
96%
99%
1%
Concurrence of System and Laboratory Reference
The percentage of concurring CGM readings and YSI reference values
were included in Table 3. For example, when CGM readings are within
81 to 120 mg/dL (4.4-6.7 mmol/L), you can expect your blood glucose
levels are within 40 to 60 mg/dL (2.2-3.3 mmol/L) 2% of time; within
60 to 80 mg/dL (3.3-4.4 mmol/L) 17% of time; within 81 to 120 mg/dL
(4.4- 6.7 mmol/L) 66% of time; within 121 to 160 mg/dL (6.7 to 8.9
mmol/L) 13% of time; and within 161 to 180 mg/dL (8.9 to 10 mmol/L) 1%
of time.
Table 3 is categorized by each range of CGM glucose ranges. This table
describes for each range of CGM glucose readings, what percentage
of paired YSI values were in the same glucose range (shaded) or in
Summary of Contents for G4 Platinum Professional
Page 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Page 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Page 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Page 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Page 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Page 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Page 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Page 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Page 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Page 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Page 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Page 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Page 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Page 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Page 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Page 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Page 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Page 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Page 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Page 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Page 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Page 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Page 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Page 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...