7.2 HOW TO CALIBRATE
You must enter the exact blood glucose value from the meter for each
calibration. Blood glucose values used for calibration must be between
40-400 mg/dL and must have been taken within the past 5 minutes.
• In real-time mode, make sure either a sensor glucose reading or a
calibration needed symbol shows at the top of the trend graph in the
status area before calibrating. In blinded mode, make sure either “- - -”
or a calibration needed symbol shows at the top of the trend graph
in the status area before calibrating.
• The sensor can be calibrated if the meter glucose is 40 mg/dL or
above. For safety reasons, if the blood glucose is low, first treat the low
blood sugar.
• Always make sure the antenna symbol
is in the upper left corner of
the trend graph before you enter blood glucose values for calibration.
• Always use the same meter to calibrate that you routinely use to
measure the blood glucose. Do not switch the meter in the middle of a
sensor session. Blood glucose meter and strip accuracy vary between
blood glucose meter brands.
• The accuracy of the blood glucose meter value used for calibration
may affect the accuracy of sensor glucose readings.
WARNING
Calibrate at least once every 12 hours. Calibrating less
often than every 12 hours might cause sensor glucose
readings to be inaccurate, and you might miss a low or
high blood glucose value.
Summary of Contents for G4 Platinum Professional
Page 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Page 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Page 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Page 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Page 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Page 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Page 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Page 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Page 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Page 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Page 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Page 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Page 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Page 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Page 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Page 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Page 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Page 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Page 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Page 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Page 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Page 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Page 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Page 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...