16
For office use only. Don’t give reader to patient.
Electromagnetic Immunity and Emissions Declaration
and Guidance
The Reader is intended for use in the electromagnetic environment
specified in the next table . The customer or the user of the Reader
should ensure that it is used in such an environment .
Immunity Test
Reader Compliance Level
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field (50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power supply lines
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
0% 230V for 1 cycle
0% 230V for 0 .5 cycle at 8 phase angles
70% 230V (30% dip in 230V) for 25 cycles
0% 230V for 250 cycles
Conducted Fields
Disturbance
IEC 61000-4-6
6 Vrms
150 kHz to 80 MHz
Radiated Fields
Disturbance
IEC 61000-4-3
10 V/m at 80 MHz to 2700 MHz (AM
Modulation)
Radiated and
Conducted Fields
Aircraft use
FAA RTCA /DO-160 edition G Section 20
Category T .
Can be used on aircraft according to the
directions provided by the operator of
the aircraft