DEZEGA
Р-70i
9
Oxygen cylinder
Make sure that the oxygen cylinder valve is closed and the connections are not contaminated with
oil or grease.
•
Manually loosen the connection between the cylinder valve and the oxygen distribution unit
and disconnect them.
•
Remove the edge of the strapping tape on cylinder left side from the pin and loosen it.
•
Hold the cylinder by valve and take it out by moving aside from oxygen distribution unit. With
the help of another hand move the nut of oxygen distribution unit away from the valve by
squeezing the spring.
CAUTION!
Do not use tools to disconnect the cylinder. Only manual disconnection of the oxygen cylinder is
allowed.
Breathing system
The breathing system contains breathing hoses, inhalation and exhalation valves, a regenerative
cartridge, a cooler, and breathing bags.
•
Disconnect crosstop’s lever and frame, by lifting the catch on the lever.