ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc.
Page 108
The FPG8601 readings may also become noisy if the piston is no longer well centered in the
cylinder. This may occur when installing the piston-cylinder or due to back pressure or lack
of lubricating pressure. When this occurs, the piston-cylinder needs to be manipulated to
reestablish the gas film between the piston-cylinder that lubricates them so the piston is
perfectly mobile. When FPG8601 pressure readings are noisy, unless very obvious signs of
contamination are present, always try piston-cylinder manipulation to establish normal
behavior before disassembling and cleaning the piston-cylinder. See Section 6.3.7 for the
procedure to reestablish piston mobility.
Failure to follow each of the FPG8601 piston-cylinder disassembly/assembly steps can
cause permanent damage to essential parts of the FPG8601.
1. Upper mounting post
4
Gimbal ring
2. Piston retaining nut
5. Cylinder
3. Indexing block with 3 mm indexing pin
6. Piston
Figure 46.
FPG8601 piston-cylinder assembly in mounting post
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...