7. MAINTENANCE OF TRACEABILITY AND RECALIBRATION
Page 155
© 2005 DH Instruments, Inc.
To remove the PG7601 reference vacuum CDG assembly and power supply/read
out, proceed as follows:
Prepare the ADCS-601 for removal of the CDG:
Vent the ADCS-601 system by pressing the
[VENT]
button on the
ADCS
Run Screen
(see Section 4.2.3). Wait for the operation to complete.
Put the system into gauge mode by pressing the
[MODE]
button on the
ADCS Run Screen
. Wait for the operation to complete.
Press
the
[Pause]
button on the
ADCS Main Tool Bar
to suspend remote
communication within the ADCS-601 system (see Section 4.3.3).
Press
the
[ESC]
key on the PG7000 Terminal.
Remove the CDG:
Disconnect the CDG isolation valve (valve 3) 1/8 in. pneumatic actuation line
(
V9
) using the quick connector on the valve.
Disconnect the electrical cable on the top of the CDG.
Break the KF40 fitting on the AMH-38 vacuum chamber or the KF25 fitting on
the adaptor. DO NOT BREAK ANY FITTINGS ON THE CDG SIDE OF THE
VALVE.
Remove the CDG supply/readout:
Remove the Control Cabinet main rear panel to access the inside of the cabinet.
Reach inside the cabinet to access the panel mount retainer screws on the
CDG supply/readout. Loosen the screws and remove the supply/readout.
When shipping the CDG for calibration include the CDG, the supply/readout and
the cable that connects the CDG to the supply read/out.
When the CDG and its power supply/readout are received from recalibration,
reinstall them onto the PG7601 and into the ADCS-601 control cabinet.
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...