3. INSTALLATION
Page 51
© 2005 DH Instruments, Inc.
When
<Leading characters to strip>
is displayed, press
[ENT]
to accept the
value of
<0>
.
3.4.2.3
EDIT AND SELECT PG7601 SETUP FILE
On the PG7601 Terminal front panel, select
[SETUP]
,
<3edit>
.
When
<Edit SETUP file:>
is displayed, press
[ENT]
to accept
<#2 >
.
Then select
<7vac>
,
<4COM2>
. Press
[ESC]
and in response to <
Save as
SETUP file:>
press
[ENT]
to accept
<#2 >
. This returns operation to the MAIN
RUN screen.
Select
[SETUP]
,
<1select>
.
When
<Use SETUP file:>
is displayed, enter
<2>
.
<Initializing extern vac sensor>
should be displayed briefly. If a failure
message is displayed, check COM ports (see Section
Error! Reference source
not found.
)
3.4.3
PG7601
3.4.3.1
POWER UP
In normal ADCS-601 operation, the PG7601 is controlled remotely by
ADCS
Tools
software. For power up and initialization, the PG7601’s PG7000 Terminal
keypad and display are used.
Turn the PG7601 power ON by pressing the power ON/OFF switch on the rear
panel of the PG7000 Terminal. Observe the display as the terminal connects
with the PG7601 Platform, tests, initializes and goes to the main run screen.
If
<….Searching…..>
displays for more than 5 seconds, the communications
between the PG7000 and the PG Terminal are failing. Check that the PG7601 to
PG Terminal cable is properly installed.
If PG7601 fails to reach the main run screen, service may be required. Record
the sequence of operations and displays observed and contact a
DHI
Authorized
Service Provider (see Table 26).
3.4.3.2
INITIALIZATION
VERIFY THE PISTON-CYLINDER MODULE, MASS SET AND MASS
LOADING BELL
PG7601 uses stored piston-cylinder and mass set metrological information to
calculate the reference pressures it defines (see the PG7000 Operation and
Maintenance Manual for more detail). For the pressure values to be correct, the
stored metrological information on the piston-cylinder and mass set must be
correct. Before using PG7601 for accurate pressure definition, the validity of the
stored information should be verified. This consists of comparing the piston-
cylinder and mass set information stored in PG7601 to the information in the
current piston-cylinder and mass set calibration reports.
With a new ADCS-601 system, the piston-cylinder module and mass set
information have already been loaded into the PG7601.
To verify the PG7601 piston-cylinder and mass set information, use the viewing
capabilities accessed by pressing
[SPECIAL]
,
<1PC/MS>
on the PG7000 Terminal.
Compare the information contained in the PG7601 piston-cylinder and mass set files
to the information on the piston-cylinder and mass set calibration reports.
VERIFY SETUP FILE
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...