3.9.5.5
<5LEVEL> (SECURITY)
PURPOSE
To set user protection levels to restrict access to certain functions and to edit the
password required for changing user levels.
PRINCIPLE
RPM4-E-DWT’s front panel user interface provides the means to access all
RPM4-E-DWT user defined data, settings and functions including calibration data.
Inadvertent, uninformed or unauthorized altering or deleting of data, settings and
functions could require extensive reconfiguration by the user and might cause invalid
readings and behavior. For these reasons, depending upon the application in which
RPM4-E-DWT is being used, it may be desirable to restrict access to some
functions. The user level function provides a means of restricting access to certain
functions. Four different levels of security are available.
Access to changing security levels can be left open, or be protected by a
password.
Security Levels
The security levels are structured to support typical operating environments as
follows:
None
This level is intended for use only by the system manager and/or
calibration facility. It allows access and editing in all areas including
critical metrological information.
Low
Low security is designed to protect the specific metrological
information and SYSTEM DIAGNOSTIC AND MAINTENANCE
functions of the system against accidental alteration. It is intended for an
advanced operator performing many different tasks. Low security is
the default user level setting.
Medium
Medium security is designed to protect specific metrological
information in the system and to assure that the RPM4-E-DWT is
operated using consistent operational parameters.
High
High security is designed to protect all operating parameters. It is
intended to minimize operator choices, for example to perform
repeated identical calibrations under consistent conditions.
RPM4-E-DWT is delivered with the security level set at low to avoid inadvertent
altering of critical internal settings but with access to changing security levels
unrestricted. It is recommended that the low security level be maintained at all
times. If there is a risk of unauthorized changing of the security level, changing
authority should be password protected (see
OPERATION of this section).
The security levels are structured to support typical levels of operation as shown
in Table 5. Specifically, the security levels prevent execution of the functions
accessed by the key strokes marked by “•”: