3. OPERATION
Page 67
© 2008 DH Instruments, a Fluke Company
Table 8.
Reset – Sets
RESET
RESULT
SEE SECTION
[UNIT]
Pressure unit of measure to first of the six available
3.7.3
[RANGE]
Hi Q-RPT default range
3.7.2
[HEAD]
0 in height and Oil medium
3.7.8, 3.9.3
Stability Limit
0.01 % of Hi Q-RPT default range
3.8.3
Upper Limit
Hi Q-RPT default range default value
3.8.4
Resolution
0.01 % FS of Hi Q-RPT default range
3.8.2
AutoZ
AutoZ ON for all Q-RPTs, Z
offset
value not affected
3.9.1
AutoZ
P
atm,0
set to 101.325 kPa a
3.9.1
Leak Check
15 second run time. Clear logged results
3.7.6
Screen Saver
10 minutes to activation
3.9.5.1
Key Sounds
Medium tone valid key press sound
3.9.5.2
Lo Vnt
Automatic
None
ReadRate Automatic
3.9.7.2
3.9.9.2
<2 UNITS>
PURPOSE
Sets the pressure units available under the UNIT function to default selections
depending (see Section 3.7.3).
Sets the user defined unit to 1.000/Pa (see Section 3.9.6).
Sets the reference temperature for inWa unit to 20°C.
3.9.9.3
<3AT
EST
>
PURPOSE
•
Resets Auto Tests to default characteristics (see Section 3.7.11).
•
Clears File AutoTest definition files (see Section 3.8.5.2).
•
Clears AutoTest results data logs (see Section 3.8.5.1).
3.9.9.4
<4 CAL>
PURPOSE
The Reset - Cal function will reset the Q-RPT and barometer calibration
coefficients and reset AutoZ values to zero. This will change the RPM4-E-DWT
calibration and could cause it to make out of tolerance measurements.
Clears all user values affecting the calibration of Q-RPTs and the on-board
barometer. The Reset – Cal resets are itemized in Table 9.