background image

    

molstic™, hi flow, molbloc™ Mounting System Instruction Sheet

molstic™, hi flow, molbloc™ Mounting System Instruction Sheet

molstic™, hi flow, molbloc™ Mounting System Instruction Sheet

molstic™, hi flow, molbloc™ Mounting System Instruction Sheet

 

 

 

molbloc, molbox and molstic are trademarks of 

DH Instruments, Inc.

 

Nupro is a registered trademark of NUPRO Co.  NUPRO is a Swagelok company. 

SWG and VCR are registered trademarks of The Swagelok Companies. 

Document 560040a 

010626 

DH Instruments, Inc. 

4765 East Beautiful Lane 
Phoenix  AZ  85044-5318 
USA 

 

 

Calibration Solutions 
for Pressure 
and Flow™ 

Tel 602.431.9100 

Fax 602.431.9559 

dhi@dhinstruments.com 

www.dhinstruments.com 

 

molbloc INSTALLATION 

Loosen the interconnecting nipple clamp (3) so that the nipple can move freely in its rubber grommet.  
Move the nipple away from the molbloc cradles to allow maximum room for molbloc insertion. 

#

#

#

#

 

Place the molbloc in the molstic molbloc cradles (be sure flow orientation is correct).  Check that no debris 
is under molbloc. 

$

$

$

$

 

Connect and tighten the upstream (valve side) molbloc VCR connection using a soft O-ring and following 
the procedure provided in Document 560009 (Recommendations for molbloc Installation) supplied with the 
molbloc. 

%

%

%

%

 

Pull the interconnecting nipple towards the molbloc downstream connection.  Connect and tighten the 
downstream molbloc VCR connection using a soft O-ring and following the procedure provided in 
Document 560009 (Recommendations for molbloc Installation) supplied with the molbloc. 

&

 

Tighten the interconnecting nipple clamp (3) until the nipple is secure (5 N/m max). 

'

'

'

'

 

Connect the molbloc to the molbox and leak check the system. 

(

(

(

 

 

 

To remove a molbloc from the molstic, reverse the procedure.  Break the downstream molbloc VCR 
connector and pull the interconnecting nipple away from the bloc.  Break the upstream molbloc 
VCR connector.  Remove the molbloc. 

DUT PLATFORM INSTALLATION AND OPERATION 

Insert socket cap screws (1) through two parallel mounting holes in the molstic 
base corresponding to the desired DUT platform position. 

#

#

#

#

 

Thread thumbwheels (2) onto screws (1) to about half way up the screw. 

$

$

$

$

 

Using a 4 mm Allen wrench, tighten the screws (1) into the DUT platform (3).  
Tighten until snug, do NOT overtighten. 

%

%

%

%

 

Rotate thumbwheels (2) to position the DUT platform at the desired height so that the DUT fitting aligns 
with the molstic downstream fitting.

 

MAINTENANCE AND RECOMMENDATIONS 

Filter:

  The molstic filter is a sealed 2 micron pleated mesh filter (NUPRO SS-4FW-VCR-2).  It can be 

cleaned by backflushing.  To backflush, it must be removed from the molstic.  When removing, cleaning 
and installing the filter, the metallic VCR gaskets (NUPRO SS-4-VCR-2-GR) must be replaced and all 
manufacturers’ recommendations followed. 

)

)

)

)

 

Handling: 

 molstic should not be picked up or carried by its components.  Always handle the molstic by 

its base. 

Q U IC K   C O N N E C T
S T E M   W IT H   ¼  IN .
S W AG EL O K  F IT T IN G

Q U IC K
C O N N E C TO R

½   IN . 
VC R

Reviews: