PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 96
LOCAL
Purpose
Returns control to the PPC2+ front panel.
Command
“LOCAL”
Classic “LOCAL”
Remark
The REMOTE program message can lock the front panel out completely. The user
can return to local operation by sending the LOCAL program message, sending the
IEEE-488 ‘GTL’ command (if in enhanced format), or by cycling PPC2+ power.
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL”
“LOCAL” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“LOCAL”
“LOCAL”
See
Also “REMOTE”
MEM
Purpose
Read the status from the power-up memory test.
Query “MEM?”
Classic “MEM”
Remarks
The PPC2+ system memory stores the user settings (units, resolution, generation
settings) and retains them when the unit is OFF. On power-up, this memory is
checked. If this memory is corrupted, all user settings are reset to default (as if the
“RESET” program message was executed), and the MEM status will be set to reflect
this.
Example
(enhanced)
Query sent:
Reply
“MEM?”
“0” PPC2+ data corrupted and was set to factory defaults
“1” The memory was found to be OK on power-up
Example
(classic)
Sent:
Reply
“MEM”
“MEM=0” PPC2+ data corrupted and was set to factory defaults
“MEM=1” The memory was found to be OK on power-up
See Also
Section 3.4.4.5, “RESET” program message
MODE
Purpose
Read or set the control mode.
Command
Query
“MODE
n”
“MODE?”
Classic “MODE=
n”
“MODE”
Arguments
n:
’0’ for static pressure control.
’1’ for dynamic pressure control.
Remarks
The method which the PPC2+ controls pressure is selected with the MODE program
message. When the control mode is set, control parameters go to default
parameters for that range. The mode setting is range dependent.
Example
(enhanced)
Cmd sent:
Query reply:
“MODE 1”
“1”
Example
(classic)
Sent:
Reply:
“MODE=1”
“MODE=1”
Errors
ERR# 6
The argument is invalid.
See Also
Sections 3.1.2.3, 3.1.2.4, 3.2.3, 3.2.6
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...