PPC2+™ Operation and Maintenance Manual
Page 99
©2001 DH Instruments, Inc.
PCALn:HI and PCALn:LO
Purpose
Read or set the user RPT calibration information.
Command
Query
“PCAL
n:HI adder, mult, CalDate”
“PCAL
n:LO adder, mult, CalDate”
“PCAL
n:HI?”
“PCAL
n:LO?”
Classic “PCAL
n:HI =adder, mult, CalDate”
“PCAL
n:LO =adder, mult, CalDate”
“PCAL
n:HI”
“PCAL
n:LO”
Defaults “PCAL
n:HI = 0.0, 1.0, 800101”
“PCAL
n:LO = 0.0, 1.0, 800101”
Arguments
N:
The RPT range. ‘1’ for the low, ‘2’ for the medium ‘3’ for the high
range of the selected RPT (Hi or Lo).
Adder:
The RPT calibration adder (PA) from –10 000 to 10 000 Pa.
Mult:
The RPT calibration multiplier (PM) from 0.9 to 1.1.
CalDate:
The date of the calibration in the format “YYMMDD”.
Remarks
The user defined pressure calibration information for the specified RPT (HI or LO
pressure) and RPT range (1, 2, or 3) can be accessed with this program message
Using this program message will overwrite the current calibration coefficients, so
caution must be used. Changes made using this program message will take affect
immediately.
Example
(enhanced)
Cmd sent:
Query reply:
“PCAL1:LO 2.1, 1.000021, 961201”
“ 2.10 Paa, 1.000021, 961201”
Example
(classic)
Sent:
Reply:
“PCAL1:LO=2.1, 1.000021, 961201”
“ 2.10 Paa, 1.000021, 961201”
Errors
ERR# 6:
One of the arguments is out of range.
See
Also Section
5.2
PPC
This is an obsolete PPC2 Command.
Purpose
PPC2+ does not support external devices so the “PPC” command is not supported.
Will respond ERR# 9.
See None
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...