PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 100
PR
Purpose
Read the next available pressure.
Query
“PR?”
Classic “PR”
Remarks
The next available pressure value for the active RPT and range is read in the current
pressure units. The data returned also contains ready information, and the pressure
units.
The reply field is always 20 characters long. The first 3 characters of the reply are
reserved for the ready status (R or NR). The ready status is covered in the “SR”
program message. The pressure measurement value and pressure units are right
justified in this field.
After receiving this program message, the PPC2+ will reply back with the data once
a new pressure measurement cycle is complete. This can take up to 1.5 seconds.
Example
(enhanced)
Query sent:
Query reply:
“PR?”
“R 1936.72 kPaa”
Example
(classic)
Query sent:
Reply:
“PR”
“R 1936.72 kPaa”
See Also
“QPRR”, “SR”, Sections 3.1.2.3, 3.1.2.4
PRR
Purpose
Read the next available pressure, rate and ATM pressure.
Query “PRR?”
Classic “PRR”
Remarks
The next available ready condition, RPT pressure, rate of pressure change, and
ambient pressure is replied in the current pressure units. Each data field is
separated by a comma, and is returned in the following order:
ready, pressure UNITS, rate UNITS/s, atm UNITS
Here are the field descriptions:
ready:
‘R’ if the current pressure ready criteria has been met,
‘NR’ if the criteria has not been met (see the “SR” program
message).
pressure:
The measured pressure for the selected RPT and range in the
current pressure units. This is followed by the current pressure units.
rate:
The measured rate for the selected RPT and range in the current
pressure units per second. This is followed by the current pressure
units.
atm:
The ambient pressure measured by the PPC2+ on-board barometer
in the current pressure units (but always absolute). This is followed
by the current pressure units.
After receiving this program message query, the PPC2+ will reply
back with the data once a new pressure measurement cycle is
complete. This can take up to 1.5 seconds.
Example
(enhanced)
Query sent:
Query reply:
“PRR?”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
Example
(classic)
Query sent:
Reply:
“PRR”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
See Also
Sections 3.1.2.3, 3.1.2.4, 3.4.4.7, “PR”, “QPRR”, “SR”
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...