PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 104
RANGE
Purpose
Read or change the active RPT range.
Command
Query
“RANGE
n:XX”
“RANGE?”
Classic “RANGE=
n:XX”
“RANGE ”
Default
“RANGE 3,HI”
Arguments
n:
‘1’ for the low range of the selected RPT.
‘2’ for the mid range of the selected RPT.
‘3’ for the high range of the selected RPT.
XX:
‘LO’ for the low pressure RPT.
‘HI’ for the high pressure RPT or if PPC2+ has only one RPT.
Remarks
The range and active RPT must be selected before making changes to settings that
are dependent on the range. Each RPTs has 3 separate ranges. You must specify
both the range and the RPT. The system MUST BE VENTED to switch RPTs. The
query reply is in the current units unless the current units are altitude units. If the
units are meters, then the range will be specified in kPa. If the units are feet, then
the range will be specified in psi.
Example
(enhanced)
Cmd sent:
Query reply:
“RANGE 1:LO”
“30 psia”
Example
(classic)
Sent:
Reply:
“RANGE=1:LO”
“30 psia”
Errors ERR#
6:
ERR# 22:
ERR# 38:
Invalid
n or XX argument.
System must be vented for the requested operation.
The selected range is not available.
See Also
Sections 3.1.2.5, 3.2.1
RATE
Purpose
Read the next available pressure rate of change.
Query “RATE?”
Classic “RATE”
Remarks
The next available pressure rate of change in the current pressure units per second
is returned.
After receiving this program message, the PPC2+ will reply back with the data once
a new pressure measurement cycle is complete. This can take up to 1.5 seconds.
Example
(enhanced)
Query sent:
Query reply:
“RATE?”
“0.01 kPa/s”
Example
(classic)
Sent:
Reply:
“RATE”
“0.01 kPa/s”
See Also
“PRR”, “QPRR”
READY
This is an obsolete PPC2 Command.
Purpose
Has the same effect as “MODE”.
See “MODE”
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...