PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 114
VAC
Purpose
Read or set the control ref status.
Command
Query
“VAC
n”
“VAC?”
Classic “VAC=
n”
“VAC”
Default “VAC
0”
Arguments
n:
’0’ To disable auto mode and specify that the PPC2+ EXHAUST port
is open to atmosphere.
’1’ To disable auto mode and specify that the PPC2+ EXHAUST port
is connected to a vacuum source.
Remarks
The PPC2+ has an internal sensor that determines if the exhaust port is open to
atmosphere or to a vacuum. You can manually override it if desired. The query can
also be used to see the sensor status (if auto is enabled). If the auto function is
disabled, then the query returns the last VAC setting.
Example
(enhanced)
Cmd sent:
Query reply:
“VAC 1”
“0”
“1”
Example
(classic)
Sent:
Query reply:
“VAC=1”
“VAC=0”
“VAC=1”
Errors
ERR# 6:
The argument is not a ‘0’ or a ‘1’.
See Also
Section 3.3.1, “AUTOVAC”
VENT
Purpose
Read, execute or abort a vent process.
Command
Query
“VENT
n”
“VENT?”
Classic “VENT=
n”
“VENT”
Arguments
n:
’1’ to start a vent process.
’0’ to abort a vent process and close the exhaust valve.
Remarks
The PPC2+ vents by generating close to atmosphere, and then opening the exhaust
valve. This program message query will return a ‘0’ if the exhaust valve is closed, or
a ‘1’ if the exhaust valve is open.
Example
(enhanced)
Cmd sent:
Query reply:
“VENT 1”
“0” (if not vented)
“1” (if vented)
Example
(classic)
Sent:
Reply:
“VENT=1”
“VENT=0” (if not vented)
“VENT=1” (if vented)
Errors
ERR# 6:
The argument is not a ‘0’ or a ‘1’.
See Also
Section 3.1.2.2
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...