PPC2+™ Operation and Maintenance Manual
Page 5
©2001 DH Instruments, Inc.
Control Precision
1
: ± 0.001 % (with 30:1 maximum ratio between highest and lowest
range)
Normal Test Volume: Lo (max range
<
300 psi): 100 to 1 000 cc (500 cc optimal)
Hi (max range
300 psi): 0 to 500 cc (250 cc optimal)
Control Speed: Slew Rate (0 to Controller FS)
in Optimal Volume with No Control: 10 to 30 seconds
Pressure Setting (Typical Time To
Ready (<*>) Indication in
Dynamic Mode):
20 to 30 seconds (lower for
low, higher for high ranges)
Reduced by increasing hold
limit
Minimum Control Pressure: Gauge Mode:
None when EXHAUST at vacuum
Absolute Mode: 0.2 % FS of range H3, will also set zero absolute
(see Section 3.2.3.1).
Delivered Pressure Accuracy
2
:
90 day 1
year
Gauge Mode:
0.010 %
0.013 %
Absolute Mode (w/Autozero):
0.010 %
0.013 %
Absolute Mode (w/out Autozero):
0.012 %
0.018 %
1 Control Precision: Minimum useable hold limit in dynamic control mode.
2
Delivered Pressure Accuracy: Maximum deviation from the true value of pressure applied to the
device under test in dynamic control mode with default control limits and assuming measurement
accuracy as in Section 1.2.2.1. In static control mode, control errors can be eliminated making
accuracy of delivered pressure the same as measurement accuracy.
1.2.4
ANALOG INPUT SPECIFICATION
Table 3. Analog Input Specifications
MEASUREMENT
RANGE
DISPLAY
RESOLUTION
A/D
RESOLUTION
ACCURACY
(+/- % FS)
0 to 100 mV
0.001
0.014
0.05
0 to 5 V
0.001
0.7
0.03
0 to 10 V
0.001
0.14
0.03
4 to 20 mA
0.001
0.003
0.05
24 V Power Source: Four channels, each output independently can sink 250 mA at 24 V, but
maximum output of all four channels must not exceed 250 mA.
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...