Figure 5. Static Pressure Control Operation
Dynamic Control
Dynamic control mode is designed to set the pressure to the target value and control
continuously to keep pressure within the hold limit and as close to the target value
as possible. The advantage of this control mode is that the final pressure achieved
will be the same as the target value. However, unlike static control, in dynamic control
a control error will be included in the pressure achieved. The maximum value of the
control error is equal to the hold limit. The average value of the control error is
generally well inside the hold limit.
During dynamic pressure control, the hold limit is active. If the pressure goes outside
of the hold limit, a Not Ready (<
↑
> or <
↓
>) condition will occur. See Sections 3.1.2.4
and 3.2.3 for default hold limit values. See Section 3.2.6 for setting the hold limit.
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...