Recommendations for the Use of the AutoZ Function
The AutoZ function provides a powerful and easy to use tool for improving the
measurement accuracy specifications of a PPC2+ and reducing the calibration recall
interval by compensating for zero drift between full recalibrations. The following simple
recommendations will help assure that you use this feature to best advantage.
•
Always leave AutoZ ON when operating in the gauge measurement mode. The only
exception is when using an absolute RPT in gauge mode at pressures under 20 psi
(150 kPa) (see note above).
•
Always leave AutoZ ON when operating in the absolute measurement mode if
ZNATERR was set properly at calibration and the AutoZ routine using a valid
atmospheric reference has been run regularly.
•
Execute the run AutoZ routine only when a reference whose accuracy is known to be
significantly better than that of the PPC2+ RPT is available. Keep range ratios in
mind when comparing accuracy. A ± 0.01 % FS barometer is roughly 10 times more
accurate than an ±
0.01
% FS 150 psi (1 MPa) RPT range because the
RPT/barometer range ratio is 10:1.
•
Though this may not be practical and generally is not necessary, the best reference
with which to run AutoZ is a gas operated piston gauge applying atmospheric
pressure to the PPC2+ test port. The best day to day reference is a properly
calibrated, barometric range, DHI RPM1, RPM2 or RPM3 interfaced directly to the
PPC2+ COM2 port.
•
Allow the PPC2+ to stabilize at atmospheric pressure and ambient temperature for
5 to 10 minutes before running AutoZ.
m
OPERATION
>RO +_^YD P_XM^SYX KXN `KV_O] K\O \KXQO +8. WYNO QK_QO Y\ KL]YV_^O ]ZOMSPSM
1K_QO<:>] 1BBBB RK`O XY D8+>/<< KXN XY SXNOZOXNOX^ +_^YD \Y_^SXO K] ^ROc
K_^YdO\Y^\KX]ZK\OX^Vc^Y^RO_]O\aROXO`O\^RO::-S]`OX^ON
To access the PPC2+ AutoZ function press
[SPECIAL] and select <1AutoZ>. The display is:
3edit 4run
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...