Electromagnetic emissions and test levels
The product is intended to be used in the stated environments with electromagnetic levels as specified below. The
caregiver and/or user of the product assure that the product is used in such an environment.
Emission test
Standards
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Hamonic current emissions
IEC 61000-3-2
Class A
Voltage fluctuations and Flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
Acceptance criteria for the EMC pass criteria is unintentional movement above ±10mm is not allowed. The product
is intended to be used in the stated environments with electromagnetic levels as specified below. The caregiver
and/or user of the product assure that the product is used in such an environment.
Electromagnetic Compatibility (EMC)
Radiated Fields in close proximity, Immunity test and compliance.
Standard: IEC 61000-4-39
Dwell time: 3 sec
Test levels (A/m)
Modulation
Pulse modulation
Mod. Frequency (kHz)
Test Frequency
8
CW
CW
30 kHz
65
50% Duty cycle
2.1
134.2 kHz
7.5
50% Duty cycle
50
13.56 MHz