8
I F U
Using Bure Walker with Gait training Kit 56-388 - KIT and Belt MULTI
Step 1: Applying the Gait training kit on the patient
The walking training harness is to be used together with Belt MULTI in
various sizes. The Gait training kit contains 4 parts made of soft materi-
als: 2 straps with soft padding and 2 straps with fastening hooks.
•
Straps with padding are to be applied between the patient’s legs
and locked in the intended lock.
•
Straps with fastening hook hang loose until patient is secured in
Bure Rise & Go (see later step 3)
Step 2: Secure attachment modules to the Bure Rise & Go Double
•
The attachment modules (2 pcs) is pushed in from the inside on
the left and right sides of the armrests under the cushions. We
recommend that the modules are placed in the middle of the
armrests.
•
Lock the module with the supplied screw. Make sure that the
module is secured.
Step 3: Securing a patient in Bure Rise & Go Double
If there is a need to further secure the patient, hook the carabiner hooks
on the green bands (right and left side) in the fastening module.Tighten
the fastening strap so that it is as short as possible.
The walking training harness is an accessory used for increased safety
when walking in the product Bure Rise & Go. The product’s both straps
and fastening hooks minimize risks if patients suffer from dizziness or
similar issues.
B
Attaching the fastening straps:
•
Green strap is threaded through the buckles from the inside /
lower part (A) - up through the buckle & upper part (B).
•
Loose straps must be outside the tensioned strap during use
A
Standing Plate
Attach standing plate by hooks on the main frame of the walker.
Summary of Contents for SystemRoMedic Bure Rise & Go DB
Page 2: ...2 I F U...