Declaration – Electromagnetic Emissions

Declaration – Electromagnetic Immunity

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity Test

IEC 60601 Test 

Level

Compliance Level

Electromagnetic

Environment - Guidance

Electrostatic discharge (ESD) 
IEC 61000-4-2

± 8kV contact
± 15kV air

± 8kV contact
± 15kV air

The relative humidity should be 
at least 5%.

Electrical fast transient/burst 
IEC 61000-4-4

± 2 kV For mains 
supply lines
100kHz repetition 
frequency

± 2 kV For mains supply 
lines
100 kHz repetition 
frequency

Mains supply quality for the 
mains adapter should be that 
of a typical commercial and/or 
hospital environment.

Surge
IEC61000-4-5

± 2kV Line(s) to 
ground
± 1kV line(s) to line

± 2kV Line(s) to ground
± 1kV line(s) to line

Mains supply quality for the 
mains adapter should be that 
of a typical commercial and/or 
hospital environment.

Voltage dips, short interruptions 
and voltage variations on mains 
supply IEC 61000-4-11

0 % U

T

; 0,5 cycle 

At 0°, 45°, 90°, 
135°, 180°, 225°, 
270° and 315°

0 % U

T

; 0,5 cycle 

At 0°, 45°, 90°, 135°, 
180°, 225°, 270° and 
315°

Mains supply quality for the 
mains adapter should be
that of a typical commercial 
and/or hospital environment.
In the event of a mains 
interruption the VENTURI
system will automatically use 
internal battery power,
unless the battery is exhausted.

0 % U

T

; 1 

cycle 
70 % U

T

; 25/30 

cycles
Single phase: at 0°

0 % U

T

; 1 cycle 

70 % U

T

; 25/30 cycles

Single phase: at 0°

Voltage interruptions

0 % U

T

; 250/300 

cycle

0 % U

T

; 250/300 cycle

Mains frequency (50/60Hz) 
magnetic field IEC61000-4-8

30 A/m

30 A/m

Mains frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical commercial 
and/or hospital environment.

Note: U

T

 is the A.C. mains voltage prior to application of the test level.

Guidance and Manufacturer’s Declaration: Electromagnetic Emissions (IEC 60601-1-2)

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions
CISPR 11

Class B

The VENTURI systems are suitable for use in all 
establishments, including domestic establishments 
and those directly connected to the public 
low-voltage power supply network that supplies 
buildings used for domestic purposes.

Harmonics emissions
61000-3-2

Class A

Voltage fluctuations / flicker emissions
61000-3-3

Complies

Summary of Contents for TG600/08F

Page 1: ...DHG HE ALTHCAR E C O M VENTURI Compact User Manual...

Page 2: ...VENTURI COMPACT DHG HEALTHCARE COM...

Page 3: ...NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 User Selectable Functions 8 Setting Changing Power Unit Display Languages 8...

Page 4: ...Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with t...

Page 5: ...for closure by primary intention for wound management odour control exudate management etc Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabet...

Page 6: ...acuum power unit or power adapter on or near a heat source Never use the power adapter whilst placed on top of or near to material which is flammable or can be damaged by heat The supplied power adapt...

Page 7: ...nal s treating the wound must assess how to best use the system for an individual wound 1 Remove all packaging from the power unit and mains adapter NB The power unit internal battery may be mostly di...

Page 8: ...t and continue NPWT 7 Canisters should be replaced as required or weekly To change canister make sure power unit is in stand by mode if still running press and hold the UNLOCK button followed by the R...

Page 9: ...ower unit will then power off after 1 minute of inactivity If using mains power switch off by disconnecting power cable from power unit or turning off mains power VACUUM LEVEL Vacuum level can be adju...

Page 10: ...ey to select this function c To select the required language press and hold the THERAPY MODE button until the language is displayed Releasing the THERAPY MODE button will then set the display for the...

Page 11: ...s adequately or may interfere with device function Direct Healthcare Group recommends the use of TECcare CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to he...

Page 12: ...he canister is removed both of which cancels the message and sounder and returns the power unit to stand by mode Change canister as detailed on page 6 NB The Canister Full sounder can be silenced by p...

Page 13: ...put Voltage 12V Nominal Vacuum Application Continuous default therapy or intermittent Pressure Range 10 to 180mmHg 0mmHg 20mmHg Fixed Internal Battery 3 7V 10Wh Lithium Ion Rechargeable Cell IP Rating...

Page 14: ...ion Medical electrical equipment safety and essential performance IEC 60601 1 11 Home healthcare environment 8 4 Manufacturer s Guarantee The VENTURI COMPACT power unit is covered by a 24 month manufa...

Page 15: ...e source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to correct the interference by one of t...

Page 16: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the...

Page 17: ...in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cell...

Page 18: ...nual was correct at the time of going to press However Direct Healthcare Group reserves the right to modify the specification of any product without prior notice in line with a policy of continual pro...

Page 19: ...DHG HEALTHCARE COM USER MANUAL Notes 1 7...

Page 20: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: