background image

•  Use of NPWT in new borns, infants and children is not recommended.
•  Intended for home healthcare and professional healthcare facility environments.
•  The device is intended to be used with its carry bag.
•  Do not connect to any other medical device or equipment.
•  The power unit and power adapter are intended to be reusable and should be cleaned between patient use (refer to Care 

and Maintenance Section).

•  The 

VENTURI

 MINO vacuum power unit is non-serviceable.

•  Wireless equipment such as mobile phones should be kept at least 1 foot or 0.3 metres away from this equipment.
•  The above warnings, cautions and any safety considerations should be observed on a routine and regular basis, not only 

upon installation.

4. How to Apply NPWT

4.1 Getting Started

NB. A Wound Care Set must be used with the 

VENTURI

 MINO system to carry out NPWT (see page 4 for available options). 

Please refer to the ‘Wound Care Set Instructions for Use’ leaflet (supplied with Wound Care Set) for dressing application and 
setup.
CAUTION! The medical professional is responsible for using his/her best medical judgment when using this system. Prior to 
use, the medical professional(s) treating the wound must assess how to best use the system for an individual wound.
1.  Remove all packaging from the power unit and mains adapter.

NB: If the battery charge appears to be partially depleted on first use, the power unit should be used with the power 
adapter, to fully charge the battery during operation.

2.  Attach the drainage canister to the 

VENTURI

 MINO power 

unit by lining up the two parts, pushing together and twisting to 
lock (Fig. 1). Remove the bung from the connector port (this can 
be retained to cap the used canister prior to disposal). Ensure 
canister is correctly located and secured otherwise NO CANISTER 
alert will appear and power unit will not operate.

3.  Prepare and seal wound as described in Wound Care Set 

‘Instructions for Use’.

4.  Attach the portal drain to the power unit canister by pushing the 

tubing firmly and fully onto the tubing connector located on the 
top of the canister. Push the tubing down into the channel to 
assist with the routing of the tubing.

4.2 Operating the Vacuum Power Unit:

1.  If using the mains adapter, insert the smaller end (DC outlet) of the supplied power adapter cable into the side of 

the 

VENTURI

 MINO power unit, and the other end into the appropriate power outlet. The power adapter indicator 

should be illuminated.
NB. The battery will charge when the unit is connected to the power source (indicated by battery charge icon on 
the top right of the display screen) and provides automatic power back-up if the external power supply or adapter 
fails. It is recommended to use the power adapter when convenient to do so as this will ensure the battery is fully 
charged when needed. A fully discharged battery will take a number of hours to fully charge.

2.  Press the RUN/STOP/UNLOCK button to initialise and run the power unit (the operating pressure and battery charge 

status will be displayed).

3.  The default operating vacuum is 80mmHg. The vacuum level can be adjusted between a choice of 80mmHg or 

120mmHg by pressing the UP/DOWN arrow button.

D H G - H E A LT H C A R E . C O M

5

U S E R   M A N U A L

Fig. 1

Summary of Contents for TG600/14

Page 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Page 2: ...VENTURI MINO DHG HEALTHCARE COM...

Page 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Page 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Page 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Page 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Page 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Page 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Page 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Page 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Page 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Page 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Page 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Page 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Page 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Page 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: