5.2 Operation Controls

N.B. Before any operation buttons will function, the power unit must be unlocked* - press and hold the RUN/STOP/
UNLOCK button for 3 seconds until lock symbol     disappears.
The operation buttons on the face of the power unit (Fig. 2) provide the following functions.

D H G - H E A LT H C A R E . C O M

7

U S E R   M A N U A L

*RUN/STOP/UNLOCK
Press the RUN/STOP/UNLOCK button to initialise and run the power unit (the operating pressure and 
battery charge status will be shown on the display screen). To unlock the power unit press and hold 
the RUN/STOP/UNLOCK button for 3 seconds until the lock symbol    disappears from the display 
screen. To stop and switch off the power unit (once unlocked) press and hold the RUN/STOP/UNLOCK 

button until the power unit beeps. Pressing the RUN/STOP/UNLOCK button during operation will also illuminate the 
display screen for 10 seconds.

VACUUM LEVEL
The vacuum level can be adjusted between a choice of 80mmHg or 120mmHg by pressing the UP/
DOWN vacuum level arrow button. The power unit will begin operation at the default pressure of 
80mmHg. The selected vacuum level is shown on the display screen.

5.3 Battery Information

•  A fully charged battery should operate the power unit continuously for at least 24 hours.
•  Charge status is shown on the display of the power unit when it is in operation.
•  The power adapter will display a green light when delivering a charge to the 

VENTURI

 MINO power unit.

•  The battery is not serviceable and cannot be removed, changed or replaced.
•  Use only the mains adapter supplied with the system.

CHARGING THE BATTERY

 - The battery will charge when the unit is connected to the 

power source, indicated by the battery charge icon on the top right of the display screen, (Fig. 3) 
(only displayed when the power unit is switched on) and provides automatic power back-up if the 
external power supply fails. It is recommended to use the power adapter when convenient to do so 
as this will ensure the battery is fully charged when needed. A daily charge is recommended. A fully 
discharged battery will take a number of hours to fully charge and can be left to charge overnight. 
NB. A full charge is indicated by the ‘lightening bolt’ symbol only.

Fig. 2

RUN/STOP

/UNLOCK

VACUUM 

LEVEL

80

-mmHg

Fig. 3

Summary of Contents for TG600/14

Page 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Page 2: ...VENTURI MINO DHG HEALTHCARE COM...

Page 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Page 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Page 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Page 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Page 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Page 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Page 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Page 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Page 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Page 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Page 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Page 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Page 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Page 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: