9083P
Multifunction Tester
Operating Instructions
328A561 revision1.1
- 15 -
5 Using the 9083P
5.1 Power On
To turn the 9083P on simply rotate the rotary switch to
the required test type.
5.2 Battery Health Check
The 9083P will automatically perform battery health
checks periodically or when a new test type is selected.
Note: When the battery symbol is flashing all tests
will be inhibited and the batteries should be
replaced as described in section 8.4.
5.3
Continuity Tests
Always ensure that the circuit under test is
electrically isolated before performing continuity
tests.
If the test probes are connected across a
voltage of >30V then the 9083P will
automatically display the voltage between the
probes, the warning buzzer will sound and the
TEST key is inhibited.
Rotate the rotary switch until the
Ω
test is selected.
When the continuity test is selected, the 9083P will
display the available test options for 1 second; Buzzer,
Test Lock, Lead Zero. If the Buzzer was previously
enabled then the icon will remain highlighted. The Test
Lock and Lead Zero must be enabled each time the
Continuity test is selected.
Functions keys F1-F4 have are used to select the
options below:
F1
F2
F3
F4
Buzzer
Test Lock
ZERO
Not used
Buzzer (F1)
When enabled, the Buzzer will sound when the
continuity measurement is less than 20 ohms.