REPLACING THE FUSES
1. Disconnect the test leads from the meter and any item under test.
2. Open the fuse cover by loosening the screw on the cover using a
Phillips head screwdriver.
3. Remove the old fuse from its holder by gently pulling it out.
4. Install the new fuse into the holder.
5. Always use a fuse of the proper size and value (0.2A/250V fast
blow for the 200mA range, 10A/250V fast blow for the 10A
range).
6. Put the fuse cover back in place. Insert the screw and tighten it
securely.
WARNING:
To avoid electric shock, disconnect the test leads
from any source of voltage before removing the fuse cover.
WARNING:
To avoid electric shock, do not operate your meter
until the fuse cover is in place and fastened securely.