DOUBLE HEAD PERISTALTIC PUMP

 USER MANUAL

DOUBLE HEAD PERISTALTIC PUMP

Cat. No. B02030001

 (TWI-PUMP)

Version 6

 I 08.11.10

Summary of Contents for B02030001 (TWI-pump)

Page 1: ...de Inc North America Phone 1 862 209 4680 Fax 1 862 209 4681 Mail orders na diagenode com DIAGENODE DOUBLE HEAD PERISTALTIC PUMP USER MANUAL DOUBLE HEAD PERISTALTIC PUMP Cat No B02030001 TWI PUMP Version 6 I 08 11 10 ...

Page 2: ...diagenode com orders diagenode com Diagenode Inc USA NORTH AMERICA 400 Morris Avenue Suite 101 Denville NJ 07834 USA Tel 1 862 209 4680 Fax 1 862 209 4681 techsupport na diagenode com orders na diagenode com For a complete listing of Diagenode s international distributors visit http www diagenode com en support distributors php For the rest of the world please contact Diagenode s a ...

Page 3: ... 4 Equipment Components 5 Technical Specifications 6 Environmental Conditions 6 Equipment installation 7 Location requirements 7 Global scheme 7 Cool Water Device Installation 8 Pump Device Installation 9 Instructions for use 14 Identification of operating controls 14 Procedure for setting 14 Equipment maintenance 15 Maintenance and inspection 15 Storage 15 Technical Assistance 15 Ordering Informa...

Page 4: ...de s charge including any costs required to return the repaired or replacement product to you This warranty covers only manufacturing defects and does not cover any damage caused by misuse non respect of recommendations described in this manual neglect accidents abrasion exposure to extreme temperatures solvents acids We strongly recommend that maintenances or repairs on your product are performed...

Page 5: ... Double head peristaltic pump 2 3 Plastic Nozzle Connector 2x2 4 Tie Wraps 10 4 Small Tie Wraps 4 5 Sturdy Tubing Inner diameter 6 35 mm Outer diameter 9 25 mm PharMed BPT Life time expected 1 000 hours 2 pieces 6 Silicon Tubing Inner diameter 8 mm Outer diameter 12 mm Transparent 6 m 7 Tube Holders Composed of two parts one rigid plastic side one gluing side 10 8 Black Plastic Connector 2 1 2 3 6...

Page 6: ... In case of infection the entire water circulation loop tubing can be sterilised by autoclave or ethylene oxide as well as by gamma radiation Environmental Conditions Devices are designed to be safe under the following conditions indoor use altitude up to 2 000 m operating external temperature 0 C to 40 C fluid temperature 4 C to 50 C maximum relative humidity 80 for temperatures up to 31 C decrea...

Page 7: ...starting installation work turn the main switch off beyond power connection and secure the unit against being re energized No special tools required Installation shown below is performed on a NESLAB RTE Bath Circulator Location Requirements Never install this equipment in a place where Environmental Conditions mentioned above are infringed 3 meters square needed to set up the whole equipment Dista...

Page 8: ...HEAD PERISTALTIC PUMP USER MANUAL Europe Diagenode sa CHU Tour GIGA B34 3e étage Avenue de l Hôpital 1 4000 Liège Sart Tilman Belgium Phone 32 4 364 20 50 Mail info diagenode com b Connection to Bioruptor Twin two waterbathes ...

Page 9: ... manual from Cool Device and perform as mentioned Note Some kind of Water Reservoir Chillers as so called Cold Bath Circulators include pumps they might not be appropriate for your use Switch them off if possible If there is no switch off selection water flow has to be turned into a continuous loop by any means See pictures below as examples Water outwards Nozzles Use spare Tubing 8mm I D x 12mm O...

Page 10: ...ld flow as shown on pictures Warm flow pipe must discharge close to the cooling coil from the chiller water reservoir Cold flow must draw water far away from the warm flow pipe discharge This is to prevent the cold flow pipe drawing warm water At least two plastic holders per pipe are required to fix appropriately the pipe because it moves constantly due to peristaltic pump head movement Before gl...

Page 11: ...ou want to integrate the pump Then connect each silicon tube ending to the appropriate sturdy tube ending Watch direction of flow each pump has a drawing side and a discharging side Heads spin clockwise which means left nozzles are the discharging side and right ones are the drawing side Have a look on the global scheme page 7 to make sure Place plastic pipe holders to fix pipes along Concerning t...

Page 12: ...ter The tubing can be cut gradually with a pair of scissors to allow a progressively higher level of water The required level is determined by the blue line water level in the tank Connect a second piece of silicone tube with the lower part of the connecting tool which is connected to cold flow Cut a piece long enough to allow the water flow going into the water For the connection of a second soni...

Page 13: ... Phone 1 862 209 4680 Fax 1 862 209 4681 Mail orders na diagenode com DIAGENODE DOUBLE HEAD PERISTALTIC PUMP USER MANUAL Then connect the pipes to the Bioruptor waterbath Insert the silicon 6 x 9 mm pipes in the black plastic connectors ...

Page 14: ...ication bath does not allow water movement upper than 500 ml min Income water flow cold flow should be adjusted around 450 ml ml The back flow warm flow must be at least the same of the income flow Note The drawing flow cannot be lower than the discharging flow otherwise the volume of water in bath would increase upper to the limit On the other hand the warm flow is able to be higher than the cold...

Page 15: ...f you notice any abnormalities with the speed of flow check it using a graduated beaker and a timer Fuse replacement Unlock the fuse drawer with a narrow blade screwdriver Info F 1A L 250 Vac Fuse reference Storage Pour water out of tubing and remove it off of pump mechanism when pumps will not be used for a long time The tubing wall may be damaged if the rollers compress the same part of the tubi...

Page 16: ... For rest of the world please contact Diagenode s a Ordering information 2010 Diagenode Inc All rights reserved The content of this document cannot be reproduced without prior permission of the authors Bioruptor and IP Star are registered trademarks of Diagenode Inc Description Cat No Format Bioruptor Twin sonication device with 1 5 ml tube holder new soundproof box UCD 400 TM 1 unit Bioruptor Twi...